The H230D helium mass spectrometer leak detector uses a 15m^3/h vortex dry pump as the front pump. It adopts an aviation specific trolley, designed for preventing collision and shock . as well as unique and convenient movement and self-locking functions. It has good passability, high sensitivity, and convenient operation. The modular design is easy to maintain. The core components are imported German brands, ensuring long-term stable and professional operation.
Helium mass apectrometer Ieak detector H230D
1.H230 adopts a split design and can move and lock the aviation trolley. making it easy to move and fix;
2.The front pump adopts a dry pump to avoid oil and gas emissions polluting the working environment;
3.Adopting user-friendly operating software. with simple icons, intuitive interface. and easy setup and operation;
4.Sniffer/negative pressure detection mode. can be freely set and switched;
5.The mass spectrometry adopts a 180 ° non-uniform magnetic field and fully automated operation. with high testing accuracy;
6.The filament adopts double iridium filament. and switching freely;
7.Automatically scan for helium peaks. Automatic calibration. Automatic range switching;
8.R232, 485, USB interface for easy system integration;
index | parameter | |
Minimum detectable leakage rate | vacuum | 5*10^-13Pam^3/s |
sniffer | 5*10^-9Pam^3/s | |
port pressure | Gross | 1500Pa |
Fine | 200Pa | |
Ultra | 40Pa | |
Leak detection mode | vacuum/sniffer/PPM | |
Start time | ≤2Min | |
response time | ≤0.5s | |
Range | 1×10^-1~1×10^-13Pa·m^3/s | |
Pressure unit | Pa、atm、mbar | |
Leakage rate unit | Pa·m^3/s、mbarL/s、atm.cc/s、ppm | |
pumping speed | 15m^3/h(4L/s) | |
port type | DN 25 ISO-KF(KF25) | |
I/O | R232、485、USB | |
power supply | AC 220V±10%50HZ | |
power | 1000W | |
dimensions(L*W*H) | 845mm*465mm*988mm | |
вес | 80kg | |
temperature | 5-40℃ |
parts | Brand/Model |
Molecular pump(optional) | Leybold/Turbovac90 |
Pfeiffer/SplitFlow80 | |
vacuum pump | Leybold/PLUS15(speed:15m^3/h) |
solenoid valve | Burkert/0256 |
amplifier | AD/LMC6041IN |
vacuometer | inficon/PSG500 |
standard leak | HLing/vcl-1(With certificate) |
ion source | HLing/Double iridium filament(coated with yttrium oxide) |