Helium mass spectrometer leak detector H230

1.H230 adopts a split design and can move and lock the aviation trolley. making it easy to move and fix;

2.Adopting user-friendly operating software. with simple icons, intuitive interface. and easy setup and operation;

3.Sniffer/negative pressure detection mode. can be freely set and switched;

4.The mass spectrometry adopts a 180 ° non-uniform magnetic field and fully automated operation. with high testing accuracy;

5.The filament adopts double iridium filament. and switching freely;

6. Automatically scan for helium peaks. Automatic calibration. Automatic range switching;

7.R232, 485, USB interface for easy system integration;

indexparameter
Minimum detectable leakage ratevacuum5*10^-13Pam^3/s
sniffer5*10^-9Pam^3/s
port pressureGross1500Pa
Fine200Pa
Ultra40Pa
Leak detection modevacuum/sniffer/PPM
Start time≤2Min
response time≤0.5s
Range1×10^-1~1×10^-13Pa·m^3/s
Pressure unitPa、atm、mbar
Leakage rate unitPa·m^3/s、mbarL/s、atm.cc/s、ppm
pumping speed16m^3/h(4.4L/s)
port typeDN 25 ISO-KF(KF25)
I/OR232、485、USB
power supplyAC 220V±10%50HZ
power1000W
dimensions(L*W*H)845mm*465mm*988mm
peso85kg
temperature5-40℃
partsBrand/Model
Molecular pump(optional)Leybold/Turbovac90
Pfeiffer/SplitFlow80
vacuum pumpLeybold/D16C(speed:16m^3/h)
solenoid valveBurkert/0256
amplifierAD/LMC6041IN
vacuometerinficon/PSG500
standard leakHLing/vcl-1(With certificate)
ion sourceHLing/Double iridium filament(coated with yttrium oxide)
es_ESSpanish